| Q: |
Why
won't a word document report be created through TWin? The
following error message is received on the Microsoft Word title
bar: "Microsoft Visual Basic ambiguous name detected: TmpDDE. |
| A: |
Go
into Microsoft Word and under the menu bar select Tools. From
the drop-down menu choose Macro. From the side menu pick
Macros. Under the Macros popup dialogue box select the file
TMPDDE and click on the delete button and close out Microsoft
Word. |
| |
|
| Q: |
When
TWin application opens, the staging status at the bottom of the TWin
PSG screen is not visible. Why? |
| A: |
Right
click on the TimeLine icon on the TWin shortcut bar
and choose Expanded CAMPRS - PSG Record and Score and right
clicking on it. Select Add this set from database and
click OK. To make it the default, after selecting Add this
set from database, right click on it again and select Select
this event set as default. Note:
Close out study and reopen for the changes to take effect. |
| |
|
| Q: |
After
Archiving, why
isn't it possible to view data from a CD? The
following error message is received: Insert another
CD. |
| A: |
File
pointer is pointing to a previous CD within
the DataBase Explorer.
Change the CD-ROM to the correct one. |
| |
|
| Q: |
Why
aren't the changes made while in Recorder visible on the Reviewer? |
| A: |
Click
on the Properties icon and go to Select Montage tab. Assure
that Use Record Montage is checked. |
| |
|
| Q: |
When
trying to delete a study from within the Patient Information Explore,
why
does the following error message appear?
File (file path)
is still recording. It
cannot be moved, copied, or delete. (PEX008)
appears. |
| A: |
This
is due to the file not being completed properly. In
order to delete the file from
within Patient Explorer, first right click on the study in
the Database Explorer
and select Terminate Record. This
will give the file a status of Completed. Now it is possible
to delete the study. |
| |
|
| Q: |
Why
is Total Sleep Time (TST) larger than Total Recorded Time (TRT) in
a recording? |
| A: |
TWin
program counts samples
to figure out time (200 samples = 1 second). At present, TWin does
not support sleep reports for 400 Hz data rate, nor is it recommended
that customers use 400 Hz with TS amplifiers, especially for sleep
studies.
|
| |
|
| Q: |
When
creating a split report, why does "0" appear in the diagnostic
section while the treatment section is fine? |
| A: |
This
usually happens when Treatment none is not added at the start
of lights out in the study. |
| |
|