Genuine Grass F-BSE1 Bipolar Stimulating Electrode Care

Cleaning & Disinfecting Grass Bipolar Stimulating Electrodes
Grass Bipolar Stimulating Electrodes can be cleaned and disinfected prior to sterilizing by soaking for 15 minutes in sodium hypochlorite. Disinfection can also be achieved by immersing the electrodes in CIDEX OPA solution, or activated glutaraldehyde (Cidex, Metricide, etc.). Follow the instructions provided by the manufacturer.
Avoid iodines, oxidating agents, peroxide, potassium permanganate and phenol agents.
 
Sterilization*
Note: Sterilization efficacy is the user's responsibility. The following methods are for reference only and Grass Technologies Product Group of Astro-Med, Inc. does not guarantee that effective sterilization will result.
 
Steam Sterilization
Note: The sterilizer manufacturer's written instructions for cycle parameters should be followed. Grass Bipolar Stimulating electrodes can be steam sterilized as follows:
Gravity Displacement Cycles 10 to 15 minutes at 270°F to 275°F (133°C to 135°C), or 15 to 30 minutes at 250°F (121°C)
Prevacuum Cycles 3 to 4 minutes at 270°F to 275°F (133°C to 135°C)
Flash Sterilization

3 minutes at 270°F (133°C) for both gravity and prevacuum cycles

Note: These are minimum exposure times which may be exceeded, if necessary.
 
Ethylene Oxide (EtO) Sterilization

Ethylene Oxide (EtO) Sterilization Grass Bipolar Stimulating electrodes can be ethylene oxide (EtO) (gas) sterilized. Follow the sterilizer manufacturer's instructions for details including: cycle time, temperature, device packaging, and aeration time.

* Recommendations from AAMI Standards and Recommended Practises. Volume 1: Sterilization

 
Electrode Tip Adjustment
The electrode ball tips are factory set for an approximate separation of 5 mm. However, the silver wire used for the electrodes is pliable and may be bent and shaped to suit the particular stimulation procedure. When bending, try not to make sharp bends, especially at the point where the electrode exits the holder or it may break.

 


©2008 Grass Technologies. All Rights Reserved.
An Astro-Med, Inc. Product Group