Page 1 [2]
AURA®24 Ambulatory EEG System
SPECIFICATIONS
Amplifier   Grass Technologies AURA® system
28 referential EEG inputs (AC),
2 reference input, 2 patient ground (with PJAAD Pin Jack Adaptor)
AC Input Signal Range   Referential AC channels:
4 mV peak to peak full-scale
Input Characteristics   12 Megohm impedance, all inputs including reference
All referential AC inputs electrically isolated from earth ground
Frequency Response   0.5 Hz to 100 Hz, -3 dB
0.1 to 100 Hz recoverable with TWin software
Amplification (gain)   AC channels: 1000 ±2%
Vertical Resolution   16 bits (0.06 µV/bit)
Sampling Rate   400 samples/second/channel
CMR   >100 dB, all AC channels, at 50/60 Hz
Noise   <2 µV peak to peak referred to input
Impedance Test Mode   Referential inputs only:
30 Hz square wave; range, 1 kohms to 100 kohms, 10%
Calibration Signal   Referential inputs only:
500 µV peak to peak ±2%;
5 Hz square wave
Patient Bias Current   Maximum: <4 nA
Typically: 300 pA
Event Input   2 pushbutton patient signaling inputs
(1 on base station, 1 on amplifier for use with 1 event pushbutton)
External Battery   Capacity: 24 hours
Charging: Built-in charger (requires external power supply)
Photic   1 photic (flash unit) control line
Oximeter   1 input (SaO2/Heart Rate)
Connectivity to Host PC   Standard Ethernet, TCP/IP protocol
Power   Medical-grade Isolated Power System
Physical Size   Amplifier Unit approximately:
3.5"W x 5.9"H x 1.0"D
(8.9 cm x 14.9 cm x 2.5 cm)
Weight: 10 ozs. (280 g)
Battery Pack approximately:
3.5"W x 5.9"H x 1.0"D
(8.9 cm x 14.9 cm x 2.5 cm)
Weight: 11 ozs. (308 g)
Base Station approximately:
7"W x 7.5"H x 2.7"D
(17.8 cm x 19.1 cm x 6.9 cm)
Weight: 2.3 lbs. (1.04 kg)
CompactFlash Card approximately:
1.4"W x 1.7"H x 0.1"D
(36 mm x 43 mm x 3 mm)
Weight: 0.4 ozs. (10 g)
Regulatory   AURA System: certified to
UL60601-1, CSA22.2 No.601.1
IEC 60601-1:1988 +A1:1991 +A2:1995,
IEC 60601-1-1, IEC 60601-2-26 (1994),
IEC 60601-1-2 (2001), Marked for ETL & CE to EC MDD (93/42/EEC)
Warranty   Lifetime Limited Warranty on Grass Technologies manufactured parts (see our full warranty for details)
Optional
Software   Powerful and user-friendly TWin record, review and analysis EEG/LTM software
  SzAC Seizure/Spike Analysis software
  Panorama Digital Video/Audio software
Operating System   Windows® XP

ORDERING INFORMATION
Model AURA24-SYS   AURA®24 includes:
Grass Technologies patient-worn AURA® Amplifier (28-channel) with 2 CompactFlash cards and reader,
2 External 24-hour Battery Packs with built-in charger and power supply,
Pouch for AURA Amplifier & External Battery Pack,
Patient Event Marker pushbutton,
Base Station with a front-end sampling rate of 400 Hz, 2 high level DC inputs,
Medical-grade Isolated Power Supply,
Network Cabling Kit.
Options:
Model AURALAPTOP   AURA®24 Laptop System includes:
Latest High-performance laptop with internal CD-RW and modem, network ready, Grass Technologies TWin EEG software, Archive Managing software, TWinLOOK® software, Microsoft Office
Model AURA-TI   On-site training (2 days recommended), USA only
Accessories:
Model UPS-1   Uninterruptible Power Source, 700 VA
(115 V, 50/60 Hz)
Model UPS-1-220   Uninterruptible Power Source, 700 VA
(230 V, 50/60 Hz)
Model UPS-2   Uninterruptible Power Source, 525 VA
(115 V, 50/60 Hz)
Options & Spare Parts:
Model PJAAD   Spare Electrode Board/Pin Jack Adaptor
Model 14193102   Spare 2 GByte CompactFlash Card
Model 42745100   Spare External Battery Pack
Model 32760100   Pouch for AURA® Amplifier & External Battery Pack