Nerve & Muscle Stimulator
S88X DUAL OUTPUT SQUARE PULSE STIMULATOR

Single, Repetitive and Train Outputs,
TTL Synchronization

NEWThe S88X is a dual output, general purpose stimulator for nerve and muscle stimulation procedures. This new stimulator has touch-pad controls.

Applications extend from single cell to entire muscle stimulation. The two output channels can be operated independently or synchronized to produce complex paradigms. The outputs are non-isolated constant voltage monophasic, biphasic or twin pulses. Up to 9 sets of parameters can be saved for future use.

New features include:
• Digital controls – precise timing accuracy
• Bright display simplifies setup
• Independent or synchronous channel operation
• Monophasic and biphasic modes
• Dual train output
• High current output facilitates field stimulation
• New companion isolation units
• Software replacement via USB cable

ORDERING INFORMATION
Model S88X
 
Stimulator including manual, power cord and (2) CAB-21891 output cables for applications requiring direct output without a stimulus isolation unit
Model S88XRK
 
Rackmount brackets to mount stimulator in a 19" cabinet
Model S88XBT   Adjustable bench-top stand
to mount one stimulator
Model
CAB-21891
  Spare Output Cable, 10-foot (3 m) cable length
Model M643H05   Spare S88X manual
Options:
Model 14004150   Host Control Starter Kit to control the S88X via the computer (USB)
Model SIU-V   Constant Voltage Stimulus Isolation Unit
Model SIU-C   Constant Current Stimulus Isolation Unit
Model SIU-BI   BiPhasic Constant Current Stimulus Isolation Unit
Model PSIU6X   Photoelectric Isolation Unit
Model CEM   Channel Expansion Module
Model SRS   Stimulate/Record Switching Unit

S88X Dual Output Stimulator

SPECIFICATIONS
Train Rate
(S1 and S2)
  1/100 sec to 99 TPS
Train Duration   1 msec to 99 sec
Rate (S1 and S2)   1/100 sec to 1000 PPS
Pulse Delay
(S1 and S2)
  10 µsec to 99 sec
Pulse Duration
(S1 and S2)
  10 µsec to 99 sec
Synchronous Outputs (TTL)   S1, S2, S1 Delay, S2 Delay, Train;
Front panel BNC connectors, 50 ohm, 10 µsec pulses, can source/sink 20 mA peak
Synchronous Inputs (TTL)   Train Duration, S1 Delay, S2 Delay;
Front panel BNC connectors, 4.75 kohm pullup to 5 volts
Stimulus Output   250 mV to 24 VDC (monophasic)
±250 mV to ±12 VDC (biphasic
)
Peak Output Current   1 A
Output Protection   Yes, current limited
Output Impedance   <2 ohms
Maximum Power Out   24 W peak
AC Power   115/230 volts, 50/60 Hz, 100 watts maximum
Physical Size   19" W x 5.25" H x 12.5" D
(48.3 cm x 13.4 cm x 31.8 cm)
Weight   10 lbs. (4.5 kg)
Regulatory   Safety: EN 61010-1:2001, UL 61010A-1, CSA C22.2 No. 1010.1-92
EMC: FCC Part 15, Subpart B, Class B, EN 61326:1998 Class B
 

PSG/Sleep Diagnostics  | EEG/LTM | Research Instruments  | Accessories | Electrodes | Electrode Application

Special Offers | Used Equipment

Online Store | Ordering Information | Place An Order | Contact Us

©2008 Grass Technologies. All Rights Reserved.
An Astro-Med, Inc. Product Group