Summary of Grass Stimulator Units
Model
S88X Dual Output Analog
S88 Dual Output Analog
S48 Single Output Analog
SD9 Output Analog
Intended Use Advanced physiological research with dual output, delay, trains. No isolation Advanced physiological research with dual output, delay, trains. No isolation Advanced physiological research with delay, trains. No isolation Student Lab, most research with delay. No trains. Output isolation
Variable Parameters Frequency, delay duration, voltage independent at both S1 and S2 outputs. Common train duration, train rate. Analog Frequency, delay duration, voltage independent at both S1 and S2 outputs. Common train duration, train rate. Analog Frequency, delay, duration, voltage, train duration, train rate from single output. Analog Frequency, delay, duration and voltage. Analog
Frequency Range 0.01 to 1000 PPS  0.01 to 1000 PPS 0.01 to 1000 PPS 0.2 to 200 PPS
Pulse Delay 10 µs to 99 sec 10 µs to 10 sec 10 µs to 10 sec 0.02 to 200 msec
Pulse Duration 10 µs to 99 sec 10 µs to 10 sec 10 µs to 10 sec 0.02 to 200 msec
Voltage Range (No load voltage) 250 mV to 24 VDC
(monophasic)
±250 mV to ±12 VDC
(biphasic
) 
10 mV to 150 volts each output 10 mV to 150 volts 0.1 to 100 volts
Peak Output Current 1 A standard 450 mA standard 650 mA standard 50 mA
Output Impedance <2 ohms  100 kohm, 250 ohm or 25 ohm each output 100 kohm, 250 ohm or 25 ohm selectable 1 kohm approx.
Maximum Power Out 24 watts peak 22 watts each output 22 watts 2.5 watts
Train Rate 1 train/100 sec to 99 TPS 1 train/100 sec to 100 TPS 1 train/100 sec to 100 TPS Not available
Train Duration 1 msec to 99 sec 1 msec to 10 sec 1 msec to 10 sec -----
Isolation of Stimulus Pulse from Ground Optional Stimulus Isolation Units (SIU), 2 required Optional Stimulus Isolation Units (SIU), 2 required Optional Stimulus Isolation Unit required Built-in (RF type)
Biphasic Pulse and Polarity Reversal Yes With optional SIUs With optional SIU Yes
Synchronous Input
(Minimum positive voltage required)
TTL compatible, +5 volts TTL compatible, +5 volts TTL compatible, +5 volts TTL compatible, +5 volts
Synchronous Outputs With stim, prestim (delay) for S1 and S2 and train With stim, prestim (delay) for S1 and S2 and train With stim, prestim (delay) and train With stim onset and prestim (delay)
Stimulus Output Combinations Single, repetitive, twin pulses, trains of pulses, trains plus post train pulses. Plus train and arrhythmic pulse. Single, repetitive, twin pulses, trains of pulses, trains plus post train pulses. Plus train and arrhythmic pulse. Single, repetitive, trains of pulses, twin pulses from single output Single, repetitive and twin pulses
Trains of Pulses Yes (both outputs) Yes (both outputs) Yes Externally controlled
AC Power Required 100 watts maximum 125 watts peak 70 watts peak 30 watts peak
Physical Size 19" W x 5.25" H x 12.5" D
(48.3 cm x 13.4 cm x 31.8 cm)
17.25" W x 8.75" H x 8.5" D
(43.8 cm x 22.3 cm x 21.6 cm)
17.25" W x 5.25" H x 8.5" D
(43.8 cm x 13.3 cm x 21.6 cm)
9.5" W x 5.25" H x 5.5" D
(24.1 cm x 13.3 cm x 13.9 cm)
Weight 10 lbs. (4.5 kg) 22 lbs. (10 kg) 17 lbs. (7.72 kg) 3.7 lbs. (1.6 kg)
   
 

PSG/Sleep Diagnostics  | EEG/LTM | Research Instruments  | Accessories | Electrodes | Electrode Application

Special Offers | Used Equipment

Online Store | Ordering Information | Place An Order | Contact Us

©2008 Grass Technologies. All Rights Reserved.
An Astro-Med, Inc. Product Group

740