These reusable
pressure transducers have long been the preferred choice of physicians
and nurses throughout the world. No other transducer can match the
performance and technical superiority. |
| |
Solid-state sensor
to ensure durability |
| |
Rugged Elgilloy
diaphragm |
| |
Lower cost per
use |
| |
Patient safety
assured by isolated plastic case and isolated sensing mechanism |
| |
Withstands 10,000
volts DC |
| |
Withstands 10,000
mmHg over-pressure |
|
|
-50
to +300 mmHg range |
| APPLICATION |
| The
P23XL provides lower cost per application, and its unsurpassed accuracy
makes it ideal for use in the range of applications from CVP, PA,
PA wedge and arterial pressures to special procedures such as intercranial
monitoring.
The
P23XL is engineered and constructed to stand up to hard, repeated
use resulting in longer transducer life for reduced cost per application.
Convenience and simplicity have always been essential criteria in
product design. The P23XL is smaller than other standard-size reusable
transducers and features a rotating locking ring for fast and convenient
dome coupling.
Miniature
Model P23XL offers unparalleled convenience. Equally impressive
is the P10EZ miniature transducer. Its small, lightweight design
permits convenient, direct patient-mounting for superb accuracy
and waveform fidelity.
|
| ORDERING
INFORMATION |
| Model
P23XL |
|
Standard
Pressure Transducer |
| Model
P23XL/T9 |
|
Standard Pressure
Transducer and 1-foot (0.3 m) shielded cable for connection to Model
LP122 Amplifier |
| Model
P10EZ |
|
Miniature
Pressure Transducer |
| Model
P10EZ/T9 |
|
Miniature Pressure
Transducer and 1-foot (0.3 m) shielded cable for connection to Model
LP122 Amplifier |
| Replacement
Domes without diaphragm |
| Model
TA1010N |
|
With
Luer-Lok fittings for use with P23XL |
| Model
TA1011 |
|
With
Linden fittings for use with P23XL |
| Model
TA1002 |
|
With
Luer-Lok fittings for use with P10EZ |
| Optional
Accessories |
| Model
TA1010ND |
|
Disposable
Diaphragm Dome, without flow/flush device, with Luer-Lok fittings
for use with P23XL |
| Model
TA1017M |
|
Disposable
Diaphragm Dome, without flow/flush device, with Luer-Lok fittings
for use with P10EZ |
| Model
TA9007 |
|
Transducer
Holder holds two P23XL transducers for pole-mount applications |
Model
TAC-1 |
|
1-foot
(0.3 m) shielded cable for P23XL, P10EZ connection to: Model 12A14,
15A12, 15AX12 Amplifiers (provided with 12A14, 15A12, 15AX12 Amplifiers)
|
Model
TAC-5 |
|
1-foot
(0.3 m) shielded cable for P23XL, P10EZ connection to: Model P11T
Amplifier |
Model
TAC-9 |
|
1-foot
length (0.3 m) shielded cable for P23XL, P10EZ connection to: Model
LP122 Amplifier |
|
| SPECIFICATIONS |
|
P23XL
|
|
Pressure
range
|
|
-50
to +300 mmHg |
| Over-pressure
tolerance |
|
10,000
mmHg |
| Bridge
resistance |
|
Input
resistance: 1,000 ohms nominal |
| |
|
Output
resistance: 350 ohms nominal |
| Excitation
voltage |
|
Up
to 10 volts DC or AC maximum, up to 5 kHz |
| Zero
offset |
|
±40
mmHg, maximum |
| Sensitivity
|
|
5
µV/V/mmHg ±1% |
| Thermal
coefficent of sensitivity |
|
±0.1
%/°C |
| Thermal
coefficient of zero |
|
±0.3
mmHg/°C |
| Electrical
leakage |
|
<2
µA at 120 VAC, 60 Hz |
| Defibrillator
withstand |
|
400
watt-seconds, maximum across 50 ohm load |
| Volume
displacement |
|
0.04
mm3/100 mmHg pressure |
| Non-linearity
and |
|
±1.5
mmHg over 0 to 300 mmHg |
| hysteresis |
|
±0.1
mmHg over 0 to 10 mmHg |
| Physical
Size |
|
1.6
inches (40 mm) |
| Weight |
|
15 grams without cable |
| Cable
length |
|
12-foot
(3.7 m ) |
|
| P10EZ |
|
Pressure
range
|
|
-50
to +300 mmHg |
| Over-pressure
tolerance |
|
10,000
mmHg |
| Bridge
resistance |
|
Input
resistance: 1,000 ohms nominal |
| |
|
Output
resistance: 350 ohms nominal |
| Excitation
voltage |
|
Up
to 10 volts DC or AC maximum, up to 5 kHz |
| Zero
offset |
|
±40
mmHg, maximum |
| Sensitivity
|
|
5
µV/V/mmHg ± 1% |
| Thermal
coefficent of sensitivity |
|
±0.1
%/°C |
| Thermal
coefficient of zero |
|
±0.3
mmHg/°C |
| Electrical
leakage |
|
<2
µA at 120V AC, 60 Hz |
| Defibrillator
withstand |
|
400
watt-seconds, maximum across 50 ohm load |
| Volume
displacement |
|
0.04
mm3/100 mmHg pressure |
| Non-linearity
and |
|
±1.5
mmHg over 0 to 300 mmHg |
| hysteresis |
|
±0.1
mmHg over 0 to 10 mmHg |
| Physical
Size |
|
1.5
inches (37 mm) including dome |
| Weight |
|
5
grams without cable |
| Cable
length |
|
12-foot
(3.7 m) |
|