| AC/DC
PRESSURE FLOW/SNORE SENSOR |
|
|
| Model
14840-KIT |
The 14840 Multi-Press provides you with AC flow and snore sensitivities adjustable through digital control over 5 levels. This sensor will automatically turn itself on when breathing is detected and automatically turn off when breathing is not detected for 5 minutes. The Multi-Press has calibrated DC pressure output as well as AC flow and snore channels for optimal use in any sleep lab. When used as a DC transducer, replacing the batteries will automatically set the module back to zero. |
|
| SPECIFICATIONS |
| Pressure
Range |
|
-3 to 40 cm H2O, -0.15 V to 2 V DC Output air |
| DC Signal Output Calibration |
|
50 mV/1 cm H2O |
| AC Output Sensitivity |
|
About 1 mV for normal respiration at midrange sensitivity |
| Snore Output Sensitivity |
|
About 1 mV for normal respiration at midrange sensitivity |
| Overall Accuracy |
|
±0.7 cm H2O over 20°C to 40°C range |
| Power |
|
2 replaceable AA batteries, good for about 1000 hours |
| Physical
Size |
|
2.8” x 4.5” x 1” (73 x 116 x 24 mm) |
| Weight |
|
Approx. 2.5 oz (72 g) |
| Cable |
|
Includes three sets of leads coming from the 14840 output: |
| |
DC Air pressure output, 14” (36 cm) long terminating in miniature phone plug output |
| |
AC coupled pressure low level output, 9” (23 cm) long terminating in two SAFELEADs |
| |
Snore low level signal output, 9” (23 cm) long terminating in two SAFELEADs |
| ORDERING
INFORMATION |
| Model
14840-KIT |
|
AC/DC Pressure Flow/Snore Sensor Kit includes sensor and PS14840 Multi Press Output Cable |
| Model PS14840 |
|
Spare Multi Press Output Cable |
| Model 14805-FT |
|
CO2 Cannula, Adult Size, 5/pk,
7-foot (2.1 m) length |
| Model 14805-2-FT |
|
CO2 Cannula, Adult Size, 5/pk,
2-foot (0.6 m) length |
| Model 14804-FT |
|
CO2 Cannula, Pediatric Size, 5/pk,
7-foot (2.1 m) length |
| Model 14804-2-FT |
|
CO2 Cannula, Pediatric Size, 5/pk,
2-foot (0.6 m) length |
These
devices are for diagnostic purposes only and not for use with or
as apnea monitors.
|
|